Specification HPLC
Contact Details Tell a Colleague about this site Download Brochures in PDF Format Employment Opportunities at Cecil Training Distributors Site Map Service A Brief History of Cecil Instruments
 
IonQuest
HPLC
Spectrophotometers
Analysers
Automated Dissolutions
Reflectance
Accessories
Adept Menu
About HPLC
Systems
Isocratic Preparative
Columns
Accessories
Specification
Ordering
Ordering Systems
 
Powerstream
About Powerstream
Data Processing
Gradient Control
System Requirements

 

HPLC Pump CE 4100
Flow Rate 0.001 to 10mL/min in 0.001 steps
Flow Reproducibility   Better than 0.1%
Maximum Pressure   40MPa = 400 bar
User Interface   4 line back-lit display
Full range of keys for menu and parameter selection
Fast number entry keys
Over Pressure Cut-out   Adjustable between 3 and 40MPa
Low Pressure Cut-out   Adjustable between 0 and 40MPa
Material in contact with fluids   Ceramic inserts - biocompatible
Check valves – ruby & sapphire
Pulsation   Better than ±0.5% at 2mL/min
Size & Weight   365 x 290 x 140mm, 7.5kg.
Power Requirements   115V or 230V, 50/60Hz, 30VA
   
Dual Wavelength, Scanning UV Visible Detector CE 4200
All specifications as for the CE 4201 with the following additional specifications
Dual Wavelength Operation   Simultaneous measurements at 2 wavelengths.
Output of absorbance ratio for the 2 wavelengths
Noise, Dual Wavelength   ± 2 x 10-5 A peak to peak 230/280nm 1s time constant
Drift, Dual Wavelength   ± 1 x 10-4 A/hour at constant temperature
Wavelength Scanning   Speeds up to 400nm/s triggered manually or by peak detection
Baseline Compensation   Baseline corrected and stored to give flat baseline for scans
Scan Storage   Up to 8 scans may be stored from any chromatogram
Up to 50 scans may be stored security protected
Programmed Operation   Programs may be entered to change parameters, operating modes and select up to 6 different wavelengths in any time sequence.
Printer Output   Parallel printer port
   
UV-Visible Detector CE 4201
Optical System   Double beam with 1200 lines/mm grating monochromator
Wavelength Range   190-700nm
Wavelength Accuracy   ± 1nm
Wavelength Precision   ± 0.1nm, single wavelength mode
Optical Bandwidth   8nm approximately
Straylight   < .02%, 220nm
Lamp   Deuterium lamp with elapsed time indicator
1500 hours or 1 year warranty
Absorbance Range   0.001 to 2A full scale in 11 ranges
Time Constant   0.1, 0.2, 0.5, 1, 2, 5 and 10 seconds
Display   Four-line display of parameters, menus, prompts, wavelength, absorbance etc.
Fast Number Entry   By use of special number keys
Event Marking   From keypad or by external source
Method Storage   For up to 30 methods, security protected
Second Order Filter   Fast automatic insertion at 370nm
Noise   ± 0.35 x 10-5 A peak to peak 220nm 2 second time constant
Drift   ± 3 x 10-5 A/hour at constant temperature
Linearity   Better than 1%, 0.001 – 2A
Automatic Power-up/Off   Pre-programmed from keypad to save warm up time
Sample Flow Cell   8µL x 10mm pathlength – normal
4µL x 5mm pathlength
Wetted surfaces PEEK, Teflon and Fused Silica
Other cells available
Refractive Index Sensitivity   < ± 5 x 10-4 A, water to methanol at 350nm
Flow Rate Sensitivity   < ± 3.5 x 10-3 A, for flow change 1-2mL/min, 300nm
Outputs   Two absorbance outputs and absorbance ratio to recorder. All data available to PowerStream System Manager.
Inputs   Event, autozero
Size and Weight   365 x 290 x 140mm, 17kg
Power Requirements   115V or 230V, 50/60Hz, 50VA
   
Refractive Index Detector CE 4700
Measuring Ranges   ± 10-4 delta n
Sensitivity   8 x 10-8 delta n
Noise   8 x 10-8 delta n
Time constants   0.1, 0.2, 0.5, 1, 2, 5, 10 seconds
Cell volume   9µL
Outputs   -1V to +1V, 20 bit resolution
Size and weight   360 x 140 x 260mm, 9.5kg
Power requirements   100-250V, 50/60Hz, 6VA
   
Automatic Sample Injector CE 4800
Sampling   96 vials 1.8mL / 48 vials 5mL optional
Injection volume   Flushed loop 5-500µL dispenser controlled
Injection   Rheodyne valve, 20µL loop standard
Injections per vial   9 max.
Analysis time   99 min 59 sec max.
Needle wash   Selectable, 50µL between vials
Size and weight   300 x 283 x 440mm, 15kg
Power requirements   115V or 230V ±10%, 50/60Hz, 100VA
   
Fluorescence Detector Filter CE 4501
Light Source   10W Xenon flash lamp, frequency programmable
Detector   Photomultiplier
Measuring Ranges   1-2000 mA in 11 ranges
Time Constants   0.1, 0.2, 0.5, 1, 2, 5 seconds
Fluorescence Display   6 digit LED
Flowcell Volume   4, 8 and 24µL others available
Inputs   Autozero, event mark
Outputs   1V and 10mV
Sensitivity   Better than 0.2pg anthracene in methanol
20µL flowcell, 254 excitation, 2 second time constant, 1mL/min flow, lamp flash frequency 50Hz
Dimension and Weight   365 x 381 x 146, 13.6kg.
Power Requirement   115/220V, 50/60Hz
   
Fluorescence Detector Dual Monochromator CE 4500
Light Source   10W Xenon flash lamp, frequency programmable
Detector   Photomultiplier
Excitation Monochromator   200-600nm, 12nm optical bandwidth, f 2.5
Emission Monochromator   200-600nm, 12nm optical bandwidth, f 2.5
Wavelength Scanning   Excitation and emission independently programmable
Sensitivity   Better than 100 fg Antracine in Methanol
Programmable Events   All normal functions plus wavelength programming up to 99 steps per program; 99 programs
Input Contact Closures   Autozero, event mark
Analogue Outputs   10mV, 100mV and 1V
Dimensions and Weight   457 x 420 x 160, 17kg
Power Requirements   100/240V, 50/60Hz
   
Column Ovens CE 4600 / CE 4601
Temperature range   10°C above ambient to 120°C (CE 4600)
10°C below ambient to 80°C (CE 4601)
Time to stability from ambient   15 mins to 100°C CE 4600
30 mins to 10°C below ambient CE 4601
Injection/switching valves   Provision for one each type Rheodyne
Solvent leak detector   64 levels, user selectable
Size and weight   190 x 230 x 540mm, 8kg
Power requirements   230V, 50Hz, 550VA or 115V, 60Hz, 550VA
   
Solvent Degassers CE 4020 / CE 4040
Number of flow channels   4 (CE4040), 2 (CE 4020)
Flowrate each channel   0.2 - 5mL/min
Gas removal   To better than 1ppm at 1mL/min
Size and weight   120 x 220 x 260mm, 6kg
Power requirements   150V or 230V, 200VA
   
Diode Array Detector CE 4400
Optical System   Single beam, 512 diodes
Wavelength Range   190-700nm
Resolution   4nm (1nm/diode)
Sources   Deuterium lamp and tungsten halide lamp
Absorbance Range   0.005 – 1.5A
Wavelength Accuracy   ± 1nm
Wavelength Setting   1nm
Spectral Bandwidth   1-175nm
Short Term Noise   2 x 10-5 A peak to peak at 254nm
Drift   1 x 10-3 A/hour
Flowcells   2, 4, 8 and 10mm pathlength
Weight   22kg
   
Conductivity Detector CE 4710
Detector   A.C. synchronous detection, 10kHz frequency
Range   0.01 – 5000 µS in 12 steps
Auto Zero   Offsets up to 10,000 µS
Methods   Up to 10 user defined methods
Heater   25W, vulcanised to aluminium block
Pre-heater   Teflon tubing encased in high thermal mass housing
Cell Temperature   Set at 35ºC. User adjustable ambient to 60ºC
Temperature Precision   ± 0.05ºC
Cell Volume   5µL
Pressure Rating   400psi (2.7 MPa)
Surfaces Contacting Fluid   PEEK and Teflon
Outputs   10mV, 100mV and 1V analogue and RS232
Size and Weight   254 x 305 x 178mm (w.d.h.), 5.9kg
Power   120-240V, 50/60Hz, 100 Watts
   
Electrochemical Detector CE4720
Measuring Range   0.02 – 50nA full scale
Offset Compensation   ±200nA
Time Constant   0.5, 1, 2 and 5s
Cell Volume   1.5µL standard
Size & weight   255 x 310 x 105mm, 5kg
Measuring cell 120 x 110 x 155, 1.5kg
 