CE 7260 Aquarius |
 |
CE 3021 D Super Aurius |
| Optical System |
|
Symmetrical double beam |
|
Single beam |
| Monochromator |
|
Czerny Turner; 1200 line/mm grating |
|
Littrow, 1200 line/mm grating |
| Wavelength Range |
|
190-900nm |
|
190-1000nm |
| Wavelength Accuracy |
|
0.5nm |
|
1nm |
|
| Wavelength Reproducibility |
|
0.1nm |
| Optical Bandwidth |
|
1.8nm |
| Straylight |
|
Typically less than 0.01% at 220nm and 340nm |
|
| Stability |
|
Better than 0.0001A/hour |
|
Better than 0.001A/hour |
|
| Photometric Range |
|
- 0.3 to 3A |
|
| Photometric Accuracy |
|
±0.004A at 1A |
|
The greater of ±0.005A or 1% |
|
| Cell Program |
|
For up to 8 cells |
| Method Storage |
|
For up to 30 methods, password protected |
| Wavelength Program |
|
For up to 10 wavelengths |
| Wavelength Scanning |
|
Up to 4000nm/min |
| Computer/Printer Interfaces |
|
Bi-directional serial RS232 and parallel ports |
| Power Requirements |
|
11 0V-250V, 50/60Hz, 170W |
|
| Size and Weight |
|
635 x 410 x 210mm, 35kg |
|
480 x 340 x 205mm, 18kg |
The CE 9260 Super Aquarius spectrophotometer is a top performance
spectrophotometer equalling or exceeding the specification of the
CE 7260 in all respects.
Dissolution Tester Specification
| Number of Stirred Vessels |
 |
7 |
| Stirrer Tools |
|
Paddles, baskets and cylinders |
| Water Bath Heater |
|
1000W; with stirrer |
| Stirrer Speeds |
|
20-250rpm |
| Stirrer Speed Accuracy |
|
±1 rpm |
| Stirrer Wobble |
|
Less than ± 0.2mm |
| Bath Temperature |
|
30-45°C |
| Temperature Accuracy |
|
±0.2°C |
| Interface |
|
RS232 |
| Bath Temperature |
|
LED display |
| Stirrer Speed |
|
LED display |
| Stirrer Head Opening |
|
Pneumatically supported |
Minimum Computer Specification
- Pentium 200MHz or better
- Microsoft Windows 98 or better
- Microsoft NT 4.0 with service pack 3 or later
- 64 MB of RAM or more
- Hard disk space required : 15MB
- CD-ROM drive
- XGA or higher resolution monitor
- Monitor resolution 1024 x 768 or better
|