Specification Spectrophotometers > Single Beam > Aurius
Contact Details Tell a Colleague about this site Download Brochures in PDF Format Employment Opportunities at Cecil Training Distributors Site Map Service A Brief History of Cecil Instruments
 
IonQuest
HPLC
Spectrophotometers
Analysers
Automated Dissolutions
Reflectance
Accessories
Aurius Menu
About Aurius and SuperAurius
Six Screen Width Viewing
Scanning
Quantitation
System Functions
Sampling
Fast Program Any Assay
Performance
Data Transfer to  PC
Performance Vallidation
Accessory Range
Software Modules
Specifications
Ordering Accessories and Spares
Ordering Instruments

 

Aurius Specification

Monochromator Littrow using 1200 L/mm holographic grating
Optical Bandwidth   4nm
Wavelength Range   190-1000nm; 325-1000nm visible versions
Wavelength Accuracy   Better than ±1nm
Wavelength Reproducibility   0.1nm
Self Test and Calibration   Automatic at switch on
Straylight   Typically less than 0.01% at 220nm and 340nm
Display Screen - LCD   High resolution backlit with contrast control
Display Scrolling   Provides 6 screen widths for viewing plots etc.
Autozero   Automatic by press button
Photometric Ranges   Digital display of -0.3-3A, 0-200%T, 0-9999C
Photometric Accuracy   ±0.005A or 1% whichever is greater
Photometric Noise   Less than ±0.0002A (500nm)
Stability   Better than 0.001A/hour at constant temperature
Multi-Wavelength Analysis   2 and 3 wavelength methods.
Assay Program   Enter multi-wavelength formula for automatic computation and formatting of any assay
Curve Fitting and Editing   Linear, quadratic or cubic; up to 30 standards
Cell and Wavelength Programming   Up to 8 cells and 10 wavelengths combined
Method Storage   Up to 30 methods stored in protected memory
Absorbance Ratioing   260/280nm or two selected wavelengths
Peak Seek   Automatic search programme, baseline corrected
Reaction Kinetics   Tabulation of results, reaction plots, reprocessing
Batch Sampling   100 samples, internal control, screen display
Real Time Clock   Timed and dated reports
Computer/printer interfaces   Bi-directional serial RS232 and parallel ports
Sample Temperature   Displayed and set from keyboard
Size   480 x 340 x 205mm
Weight   16.8Kg Visible; 18.6Kg UV; + 0.9Kg with printer
Power Requirements   110-250V, 50/60Hz, 170W/120W UV/Visible versions
     

SuperAurius Additional Specification

Optical Bandwidth   1.8nm
Wavelength Scale Expansion   Selectable by keyboard entry 1-100nm/cm
Scan Speed   Selectable by keyboard entry 1-4000nm/m
Photometric Plotting Ranges   Keyboard entry of limits; or automatic optimum
Baseline Flatness   Better than ±0.002A (300nm/min) most of range
Overlayed Spectra   Spectra and derivatives with or without offset
Spectral Reprocessing   Spectra manipulated and replotted over any range
Spectral Storage   Up to 100 spectra may be stored security code protected
Derivative Spectra   1st and 2nd derivatives
Time Course Plotting   Plots may be reprocessed and stored
 
 