Specification Analysers > AquaQuest
Contact Details Tell a Colleague about this site Download Brochures in PDF Format Employment Opportunities at Cecil Training Distributors Site Map Service A Brief History of Cecil Instruments
 
IonQuest
HPLC
Spectrophotometers
Analysers
Automated Dissolutions
Reflectance
Accessories
AquaQuest Menu
About AquaQuest Water Analysis
Large Screen - Compact Printing
Operational Simplicity
Application Range
Sipette and Auto Sampling
PreProgrammed Analyses for Merck Spectroquant Tests
Software Modules
Specificatioin
Ordering Instruments
Ordering Accessories

 

Optical System Littrow monochromator using 1200L/mm holographic grating; coated optics. Accommodates 100mm cells
Optical Bandwidth   4nm, all versions
Wavelength Range   190-1000nm; 325-1000nm visible versions
Wavelength Reproducibility   Better than 0.1nm
Wavelength Accuracy   Within 1nm
Straylight   Typically 0.01% at 220nm and 340nm - UV versions
Photometric Ranges   -0.3 to 3.0A, 0-200%T, 0-9999C
Photometric Noise   Less than ±0.0002A at 500nm
Photometric Accuracy   1% or 0.005A whichever is greater
Absorbance Zero Stability   Better than ±0.001A/Hr. at 500nm
Absorbance Zero Setting   0A and 100%T automatic by press button
Self Test and Calibration   Automatic at switch on
Filter and Lamp Selection   Automatic - coupled to wavelength
Display Screen - LCD   High resolution backlit screen with contrast control, displays menus, curves, scans and all data etc.
Display Screen Scrolling   Effective viewing width of 550mm
Wavelength Selection   Keyboard entry with GO TO key
Reaction Kinetics   Tabulation of results; reaction plots, reprocessing
Cell and Wavelength Programming   Up to 8 cells and 10 wavelengths in any combination
Peak Seek   Automatic with baseline flattening
Curve and Line Fitting   Linear, quadratic or cubic, with or without intercept, for up to 30 standards with curve editing
Wavelength Scanning   Scan speeds up to 4000nm/min with the scan software
Method Storage   Up to 100 methods and 30 curves, security coded
Real Time Clock   Timed and dated reports
Sipette Sampling   Under internal control with screen display
Batch Sampling   For up to 100 samples under internal control
Sample Temperature   Displayed and set from Keyboard
Computer Interface   Bi-directional serial RS232C and parallel ports
Sample compartment size   196 x 105 x 107mm (1 x d x h)
Power Requirements   110-250V, 50/60Hz, 170W/120W UV/Visible versions
Weight   16.8Kg CE 4001, 18.6Kg CE 4002
17.7Kg CE 4003, 19.5Kg CE 4004
Size   480 x 340 x 205mm (l x d x h)
     
Cecil Instruments policy is one of continuous development. We therefore reserve the right to change specification without notice.
 
  About | Displays and Printouts | Operation | Application Range | Sampling | Merck Tests
Software Modules | Specification | Ordering Instruments | Ordering Accessories