| Sampling system |
 |
Headspace pressure, using dual concentric needle with internal
air pump |
| Capacity |
|
100 vial carousel ( CE 4800-100) |
| Capacity |
|
50 vial carousel ( CE 4800-50) |
| Injection vial |
|
2 mL, 32 x 12 mm diameter |
| Loop size |
|
10-200 µL, 20 µL standard |
| Replicate injections per vial |
|
Up to 15 (dependent on loop size) |
| Random access sampling |
|
Yes |
| Priority sampling |
|
Yes |
| Injection volume reproducibility |
|
Better than 0.2%, (typically <0.15%) |
| Carryover |
|
Better than 0.005% with active needle wash
Better than 0.03% with no needle wash |
| Sample loop flush |
|
Operator selectable |
| Wash station capacity |
|
12 mL |
| Needle wash |
|
Dynamic wash system |
| Injection valve |
|
Rheodyne |
| Local control |
|
Press buttons and backlit LCD for data menu and instructions |
| Remote control |
|
PC control using Cecil PowerStream software |
| Communications |
|
RS232 and digital |
| Ambient temperature range |
|
10°C to 35° C |
| Power requirements |
|
110-250V, 50/60Hz |
| Dimensions |
|
36 x 40 x 33 cm (l x d x h) |
| Weight |
|
12 Kg |